CeraSys Incorporated
14747 Artesia Blvd Ste 5G, La Mirada, California, United States
FEI 3005243880
FDA has 3 inspections on file for CeraSys Incorporated in La Mirada, California (FEI 3005243880), from 2008 to 2012. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection: 7 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on May 3, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Eight regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2008–2012
- 1 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation
Process validation activities and results have not been documented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Design control - no procedures
Procedures for design control have not been established.
Show 3 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Process validation activities and results have not been documented and approved.
Show 3 more citations
Design control - no procedures
Procedures for design control have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
DMR - not or inadequately maintained
A device master record has not been maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not established.
DMR device specifications
The device master record does not include or refer to the location of all and device specifications.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established.
Show 2 more citations
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Lack of or inadequate DHR procedures
Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.181Device master record3 of 3
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 820.184Device history record2 of 3
Show all 11
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005243880 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005243880. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.