FDA has 7 inspections on file for Rebuilder Medical, Inc. in Charles Town, West Virginia (FEI 3004987960), from 2008 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Jan 11, 2011. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.100(a) and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 7, 2008–2022
- 1 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 7
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
DMR device specifications
The device master record does not include or refer to the location of device software specifications.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Show 1 more citation
Design validation - software validation not performed
Validation of device software was not performed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Design control - no procedures
Procedures for design control have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Maintained
Records of complaint investigations are not maintained by the formally designated unit.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 5 more citations
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been documented in the device history record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
No release signature
Finished devices were released for distribution without signature of the individual designated to authorize such release.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established, defined, and documented.
DMR production process specifications
The device master record does not include or refer to the location of production and process specifications.
Complete files maintained
Complete complaint files are not maintained.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 820.181Device master record2 of 4
- 21 CFR 820.22Quality audit1 of 4
Show all 11
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.184Device history record1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004987960, one row per refused product line.
1 refusal on 2019-12-10 · charged NO 510(K) on 1 · no FDA sample analysis
FDA firm profile- 2019-12-10
STIMULATOR, NERVE, TRANSCUTANEOUS, FOR PAIN RELIEF84GZJ
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-3749443-1/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004987960 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004987960. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.