Skip to content

Cybernet Medical Corp.

3885 Research Park Dr, Ann Arbor, Michigan, United States

FEI 3004198411

FDA has 3 inspections on file for Cybernet Medical Corp. in Ann Arbor, Michigan (FEI 3004198411), from 2009 to 2013. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Oct 30, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.198(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2009–2013
1 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Document review, approval documentation

      The documentation of approval of documents does not include the signature of the approving official.

    2. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    3. DMR device specifications

      The device master record does not include or refer to the location of device software specifications.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

    5. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    Show 3 more citations
    1. Training records

      Personnel training is not documented.

    2. Documentation

      Acceptance activities were not maintained as part of the device history record.

    3. Complaints

      Complaint files are not adequately maintained.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not established.

    2. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include the reason why and the name of the individual responsible for the decision not to investigate.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Design validation - user needs and intended uses

      Design validation did not ensure that devices conform to defined user/patient needs and intended uses.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    Show 6 more citations
    1. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    2. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review are not established.

    4. General

      Complaint handling procedures for complaints have not been established.

    5. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    6. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004198411 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004198411. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.