Fertility Technology Resources, Inc.
4278 E Us 64 Alt, Murphy, North Carolina, United States
FEI 3003889409
FDA has 3 inspections on file for Fertility Technology Resources, Inc. in Murphy, North Carolina (FEI 3003889409), from 2008 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2008–2017
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptable supplier records, inadequate records
Records of acceptable contractors have not been adequately established.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been established.
Documented evaluation
The evaluation of potential suppliers was not documented.
Quality Audits - defined intervals
Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR required activities {see also 820.181}
Finished devices were released for distribution prior to completion of activities required in the Device Master Record.
Storage procedures to prevent mix-ups
Procedures for controlling the storage of product in storage areas and stock rooms were not complete and implemented to prevent mix-ups, damage, other adverse effects.
Production monitoring and control
During production, process parameters are not fully monitored and controlled.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not complete and implemented.
Show 8 more citations
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, and implementation of corrective and preventive actions.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization and satisfies the requirements of the quality system regulations.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Supplier oversight
The type and extent of control to be exercised over the product, services, and contractors was not clearly defined.
Personnel
Employees do not have the necessary education, background, training, and experience to perform their jobs.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Lack of or inadequate procedures for handling
Procedures have not been established, defined, documented, complete, and implemented to ensure that mix-ups, damage, or other adverse effects to product do not occur during handling.
Production processes
Production processes were not developed, complete, conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
Show all 13
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.140Handling1 of 2
- 21 CFR 820.150Storage1 of 2
- 21 CFR 820.160Distribution1 of 2
- 21 CFR 820.184Device history record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003889409 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003889409. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.