Erchonia Corporation, LLC
112 Southchase Blvd, Fountain Inn, South Carolina, United States
FEI 3003462036
FDA has 7 inspections on file for Erchonia Corporation, LLC in Fountain Inn, South Carolina (FEI 3003462036), from 2009 to 2018. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 3 citations from this inspection. They are listed below.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.30(i).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 7, 2009–2018
- 3 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 7
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsor securing investigator compliance
An investigator was not complying with the signed agreement, investigational plan and requirements of the regulations and.
No financial disclosure info in investigator agreement
A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement.
Sponsors' general responsibilities
For an investigational study, proper monitoring was not ensured and IRB review and approval were not ensured.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
In-process inspections, tests, or other verification activities and approvals were not documented.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety ActCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Documentation
Process validation activities and results have not been documented and approved.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not documented.
Training records
Employee training is not fully documented.
Design changes - Lack of or Inadequate Procedures
Procedures were not completed and followed for the validation or verification of design changes before their implementation.
Show 3 more citations
Nonconforming product, Lack of or inadequate procedures
Procedures were not complete and followed for the control of products that do not conform to specifications.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not defined, documented, complete, and implemented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 812.40General responsibilities of sponsors1 of 4
- 21 CFR 812.43Selecting investigators and monitors1 of 4
- 21 CFR 812.46Monitoring investigations1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.70Production and process controls1 of 4
Show all 12
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.198Complaint files1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003462036 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003462036. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.