Skip to content

Erchonia Corporation, LLC

112 Southchase Blvd, Fountain Inn, South Carolina, United States

FEI 3003462036

FDA has 7 inspections on file for Erchonia Corporation, LLC in Fountain Inn, South Carolina (FEI 3003462036), from 2009 to 2018. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 3 citations from this inspection. They are listed below.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.30(i).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
7, 2009–2018
3 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
4 of 7
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsor securing investigator compliance

      An investigator was not complying with the signed agreement, investigational plan and requirements of the regulations and.

    2. No financial disclosure info in investigator agreement

      A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement.

    3. Sponsors' general responsibilities

      For an investigational study, proper monitoring was not ensured and IRB review and approval were not ensured.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    3. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety ActCitations not published
  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Documentation

      Process validation activities and results have not been documented and approved.

  6. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Software validation for automated processes

      Software used as part of production has not been validated for its intended use according to an established protocol.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    3. Procedure for analysis of data sources

      The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not documented.

    4. Training records

      Employee training is not fully documented.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures were not completed and followed for the validation or verification of design changes before their implementation.

    Show 3 more citations
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures were not complete and followed for the control of products that do not conform to specifications.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not defined, documented, complete, and implemented.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003462036 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003462036. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.