Skip to content

21 CFR 812.43: Selecting investigators and monitors

FDA's inspection data lists 85 Form 483 citations of 21 CFR 812.43 at 53 facilities, from December 2008 to November 22, 2024.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
85
At 55 inspections
Facilities cited
53
Distinct FEIs
Most recent citation
Nov 22, 2024
Nanospectra Bioscience Inc
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 812.43 from the Electronic Code of Federal Regulations.

§ 812.43 Selecting investigators and monitors.

(a) Selecting investigators. A sponsor shall select investigators qualified by training and experience to investigate the device.

(b) Control of device. A sponsor shall ship investigational devices only to qualified investigators participating in the investigation.

(c) Obtaining agreements. A sponsor shall obtain from each participating investigator a signed agreement that includes:

(1) The investigator's curriculum vitae.

(2) Where applicable, a statement of the investigator's relevant experience, including the dates, location, extent, and type of experience.

(3) If the investigator was involved in an investigation or other research that was terminated, an explanation of the circumstances that led to termination.

(4) A statement of the investigator's commitment to:

(i) Conduct the investigation in accordance with the agreement, the investigational plan, this part and other applicable FDA regulations, and conditions of approval imposed by the reviewing IRB or FDA;

(ii) Supervise all testing of the device involving human subjects; and

(iii) Ensure that the requirements for obtaining informed consent are met.

(5) Sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement as required under part 54 of this chapter. The sponsor shall obtain a commitment from the clinical investigator to promptly update this information if any relevant changes occur during the course of the investigation and for 1 year following completion of the study. This information shall not be submitted in an investigational device exemption application, but shall be submitted in any marketing application involving the device.

(d) Selecting monitors. A sponsor shall select monitors qualified by training and experience to monitor the investigational study in accordance with this part and other applicable FDA regulations.

21 CFR Part 812: Investigational Device Exemptions, Subpart CRead on eCFR

Citations by year

Form 483 citations of 21 CFR 812.43 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 17 in 2012

Form 483 citations of 21 CFR 812.43 by inspection year
YearForm 483 citations
20081
20099
201015
20116
201217
20133
201414
20161
20173
20184
201911
20241

How FDA words this citation

FDA's short descriptions and citation text on these 85 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • No investigator agreement21 citations
  • No financial disclosure info in investigator agreement20 citations
  • No investigation agreement statement of commitment10 citations
  • Investigator(s) not qualified7 citations
  • Monitors not qualified7 citations
Show 3 more
  • Sponsor shipped devices to unqualified person(s)7 citations
  • No curriculum vitae in investigator agreement5 citations
  • No informed consent statement in investigator agreement3 citations

Citation text

  • “A signed investigator agreement was not obtained from each participating investigator .”21 citations
  • “A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement.”10 citations
  • “A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement and a commitment to promptly update financial disclosure information if any relevant changes occur during the investigation and for one year following completion of the study.”9 citations
  • “Investigational devices were shipped to individuals who were not qualified investigators participating in the investigation.”7 citations
Show 2 more
  • “A signed agreement that includes a statement of the investigator's commitment to conduct an investigation in accordance with the agreement, investigational plan, applicable FDA regulations, conditions of approval imposed by the reviewing IRB, and conditions of approval imposed by the FDA was not obtained from each participating investigator.”5 citations
  • “Monitors that are qualified by training and experience were not chosen to monitor the investigational study.”5 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 812.43, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Nanospectra Bioscience Inc2024-11-22 · FEI 3006363987No financial disclosure info in investigator agreement
  • TriReme Medical, LLC.2019-10-30 · FEI 3008089360No financial disclosure info in investigator agreement
  • Kennedy, Philip R.2019-10-16 · FEI 3007184242No financial disclosure info in investigator agreement
  • Tusker Medical, Inc.2019-10-08 · FEI 3012130335Sponsor shipped devices to unqualified person(s)
  • Christopher J. Hartnick, MD2019-06-28 · FEI 3015376495Monitors not qualified
  • Vivek Reddy, M.D.2019-03-07 · FEI 3015186868No investigator agreement
Show 4 more
  • Erchonia Corporation, LLC2018-12-14 · FEI 3003462036No financial disclosure info in investigator agreement
  • Phraxis, Inc.2018-06-06 · FEI 3014146191Sponsor shipped devices to unqualified person(s)
  • CAO Group, Inc.2018-05-07 · FEI 3003258642Sponsor shipped devices to unqualified person(s)
  • Premier Clinical Research LLC2017-11-29 · FEI 3011321193Monitors not qualified

Often cited at the same inspection

Of the 55 inspections where FDA cited 21 CFR 812.43, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 812.43

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 812.43; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.