21 CFR 812.25: Investigational plan
FDA's inspection data lists 19 Form 483 citations of 21 CFR 812.25 at 18 facilities, from May 2009 to May 29, 2025.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 19
- At 19 inspections
- Facilities cited
- 18
- Distinct FEIs
- Most recent citation
- May 29, 2025
- Happy Health Inc
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 812.25 from the Electronic Code of Federal Regulations.
§ 812.25 Investigational plan.
The investigational plan shall include, in the following order:
(a) Purpose. The name and intended use of the device and the objectives and duration of the investigation.
(b) Protocol. A written protocol describing the methodology to be used and an analysis of the protocol demonstrating that the investigation is scientifically sound.
(c) Risk analysis. A description and analysis of all increased risks to which subjects will be exposed by the investigation; the manner in which these risks will be minimized; a justification for the investigation; and a description of the patient population, including the number, age, sex, and condition.
(d) Description of device. A description of each important component, ingredient, property, and principle of operation of the device and of each anticipated change in the device during the course of the investigation.
(e) Monitoring procedures. The sponsor's written procedures for monitoring the investigation and the name and address of any monitor.
(f) Labeling. Copies of all labeling for the device.
(g) Consent materials. Copies of all forms and informational materials to be provided to subjects to obtain informed consent.
(h) IRB information. A list of the names, locations, and chairpersons of all IRB's that have been or will be asked to review the investigation, and a certification of any action taken by any of those IRB's with respect to the investigation.
(i) Other institutions. The name and address of each institution at which a part of the investigation may be conducted that has not been identified in paragraph (h) of this section.
(j) Additional records and reports. A description of records and reports that will be maintained on the investigation in addition to those prescribed in subpart G.
21 CFR Part 812: Investigational Device Exemptions, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 812.25 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 3 in 2012
| Year | Form 483 citations |
|---|---|
| 2009 | 2 |
| 2010 | 2 |
| 2011 | 1 |
| 2012 | 3 |
| 2014 | 3 |
| 2015 | 1 |
| 2017 | 3 |
| 2018 | 1 |
| 2019 | 2 |
| 2025 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 19 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Sponsor's lack of written monitoring procedures19 citations
Citation text
- “There are no written procedures for monitoring an investigational device study.”19 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 812.25, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Christopher J. Hartnick, MD2019-06-28 · FEI 3015376495Sponsor's lack of written monitoring procedures
- New England Research Institutes, Inc.2017-07-13 · FEI 3002138054Sponsor's lack of written monitoring procedures
- Newport Brain Research Laboratories, Inc.2017-06-14 · FEI 3013516973Sponsor's lack of written monitoring procedures
Show 4 more
- Wakai, Ronald, Ph. D., Clinical Investigator2014-09-05 · FEI 3010985478Sponsor's lack of written monitoring procedures
Often cited at the same inspection
Of the 19 inspections where FDA cited 21 CFR 812.25, the share where it also cited each of these sections.
- 21 CFR 812.40General responsibilities of sponsors10 (53%)
- 21 CFR 812.140Records9 (47%)
- 21 CFR 812.43Selecting investigators and monitors8 (42%)
- 21 CFR 812.150Reports5 (26%)
- 21 CFR 50.25Elements of informed consent4 (21%)
- 21 CFR 812.100General responsibilities of investigators3 (16%)
Ask FDAgent about 21 CFR 812.25
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 812.25; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.