Compumedics USA, Inc
5015 W W T Harris Blvd Ste E, Charlotte, North Carolina, United States
FEI 3002716534
FDA has 4 inspections on file for Compumedics USA, Inc in Charlotte, North Carolina (FEI 3002716534), from 2015 to 2021. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2015–2021
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Installer records
The person installing the device did not adequately document the inspection and test results to demonstrate proper installation.
Supplier oversight
The type and extent of control to be exercised over the product and services was not clearly defined.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Service reports
Service reports do not include the required information.
Show 5 more citations
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Document change records, maintained.
Records of changes to documents were not maintained.
Management review dates
The results and/or dates of management reviews are not documented.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate instructions
inspection instructions and test procedures have not been established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Equipment control activity documentation
Equipment calibrations, inspections and maintenance activities have not been documented.
Analyzing service report
Service reports were not analyzed following appropriate statistical methods.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 2 more citations
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for verifying that servicing meets specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Records of complaint investigation
Records of complaint investigations do not include required information.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria and maintained.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Installer records
The person installing the device did not document the inspection and test results to demonstrate proper installation.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.170Installation3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.200Servicing3 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.25Personnel1 of 4
Show all 12
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.86Acceptance status1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002716534 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002716534. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.