Fresenius Medical Care Holdings, Inc.

920 Winter St Bld 920, Waltham, Massachusetts, United States

FEI 3001451489

FDA has 8 inspections on file for Fresenius Medical Care Holdings, Inc. in Waltham, Massachusetts (FEI 3001451489), from 2010 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the 14 FOIA requests for this site have asked FDA for this inspection's 483, and six more for the 483 from a later inspection, ending Jun 3, 2026, that FDA's Data Dashboard doesn't list yet.

The most recent OAI inspection ended on Aug 2, 2023. FDA has issued 3 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
8, 2010–2024
3 OAI · 2 VAI · 3 NAI
Inspections with 483 observations shown
6 of 8
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
14 requests, latest FY2023
Warning letters
· CDRH ↗
3 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and consultants have not been adequately established.

    3. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    4. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Incomplete ADE information on ICSR

      You submitted an ICSR reporting adverse drug experience information that failed to include the adverse drug experience term(s).

    2. Late submission of 15-day report

      Not all adverse drug experiences that are both serious and unexpected have been reported to FDA within 15 calendar days of initial receipt of the information.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.

    4. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices17 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    4. Complete files maintained

      Complete complaint files are not maintained.

    5. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    Show 12 more citations
    1. Specific non-conforming product procedures

      Procedures for addressing the identification and evaluation of nonconforming product were not defined.

    2. Procedures for product review,disposition lack of/inadequate

      Procedures that describe the review and disposition process for nonconforming product were not implemented.

    3. Investigation of device failures

      Complaints involving the possible failure of a device and labeling to meet any of its specifications were not investigated where necessary.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    6. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    7. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    8. Contamination, chemical or physical change, deterioration

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product.

    9. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.

    10. General

      Complaint handling procedures for receiving complaints have not been established.

    11. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    12. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

14 requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Jun 3, 2026

    6 requests. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Aug 13, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 3, 2026FDA-FOIA-2026-8596Form 483Records released
    2. Sep 3, 2026FDA-FOIA-2026-8583Form 483· 1 of 3 sitesNo records found
    3. Aug 21, 2026FDA-FOIA-2026-8129Form 483· 1 of 4 sitesRecords released
    4. Aug 12, 2026FDA-FOIA-2026-7774Form 483· 1 of 4 sitesRecords released
    5. Jul 22, 2026FDA-FOIA-2026-7078Form 483Records released
    6. Jul 14, 2026FDA-FOIA-2026-6712Form 483Records released
  • Inspection ending Oct 16, 2024NAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 27, 2025. FDAgent holds no copy that FDA posted.

    1. Aug 28, 2025FDA-FOIA-2025-10342Form 483Records released
    2. Nov 8, 2024FDA-FOIA-2024-9925Form 483· 1 of 2 sitesClosed
  • Inspection ending Aug 2, 2023OAI

    4 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-8003Form 483Closed
    2. Oct 13, 2023FDA-FOIA-2023-9020Form 483Closed
    3. Oct 6, 2023FDA-FOIA-2023-8777Form 483Closed
    4. Oct 5, 2023FDA-FOIA-2023-8754Form 483Closed
  • Inspection date not given

    2 requests. FDA released records in answer to a request for its Form 483 on Sep 30, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 16, 2026FDA-FOIA-2026-8948Form 483Records released
    2. Sep 9, 2026FDA-FOIA-2026-8740Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 8 inspections, with how many of them cited each.

Show all 16

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001451489 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001451489. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.