Moss Tubes Inc.
1929 Us Route 9, Castleton On Hudson, New York, United States
FEI 3000717944
FDA has 5 inspections on file for Moss Tubes Inc. in Castleton On Hudson, New York (FEI 3000717944), from 2012 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2012–2026
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Sep 18, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been adequately documented.
- ISO 13485:2016 Clause 7.5.1
Production and service provision not adequate
Production and service provision was not adequately planned, carried out, monitored, and controlled to ensure that product conforms to specification.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures not documented
Procedures for timely complaint handling have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Design validation - documentation
The results of design validation, including method(s), were not documented in the design history file.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been documented and adequately approved.
Documentation
Acceptance activities were not documented and maintained as part of the device history record.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Show 1 more citation
Reporting Information Reasonably Known
An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - documentation
The dates of quality audits have not been documented.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 18, 2026. Each one asked for a Form 483.
Inspection ending Jul 23, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 23, 2026. FDAgent holds no copy that FDA posted.
- Sep 18, 2026FDA-FOIA-2026-9054Form 483Records released
- Aug 26, 2026FDA-FOIA-2026-8267Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 5
- 21 CFR 820.30Design controls2 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.100Corrective and preventive action1 of 5
- 21 CFR 820.184Device history record1 of 5
Show all 9
- 21 CFR 820.198Complaint files1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000717944 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000717944. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.