Nidek, Inc.
6601 Cascades Ct, Lewisville, Texas, United States
FEI 2936921
FDA has 7 inspections on file for Nidek, Inc. in Lewisville, Texas (FEI 2936921), from 2012 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Apr 14, 2014. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2012–2019
- 1 OAI · 2 VAI · 4 NAI
- Inspections with 483 observations shown
- 4 of 7
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 5 more citations
Training records
Personnel training is not documented.
Investigation of device failures
Complaints involving the possible failure of a device, labeling, and packaging to meet any of its specifications were not evaluated and investigated where necessary.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated based on their ability to meet specified requirements.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 2 more citations
Training records
Personnel training is not documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been maintained.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 4
- 21 CFR 806.10Reports of corrections and removals2 of 4
- 21 CFR 820.25Personnel2 of 4
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
Show all 10
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2936921 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2936921. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.