Skip to content

Medical Components, Inc dba MedComp

1499 Delp Dr, Harleysville, Pennsylvania, United States

FEI 2518902

FDA has 6 inspections on file for Medical Components, Inc dba MedComp in Harleysville, Pennsylvania (FEI 2518902), from 2009 to 2021. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2009–2021
0 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
2 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    3. Documented evaluation

      The evaluation of potential suppliers was not documented.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Reprocessing procedures lack steps to be taken

      Reprocessing procedures lack the steps to be taken to insure that reprocessed batches will conform with all established standards, specifications, and characteristics.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete.

    3. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to procedures were not implemented.

    4. Management review - Lack of or inadequate procedures

      Procedures for management review are not complete.

    5. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not implemented.

    Show 5 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed where necessary.

    2. Quality control unit review of records

      Drug product production and control records, are not reviewed by the quality control unit to determine compliance with all established, approved written procedures before a batch is released or distributed.

    3. General

      Complaint handling procedures for evaluating complaints have not been defined.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    5. Documentation

      Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, and implementation of corrective and preventive actions.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2518902, one row per refused product line.

1 refusal on 2025-04-28 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis

FDA firm profile
  • 2025-04-28

    CATHETER, HEMODIALYSIS, IMPLANTED78MSD

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast Imports799-6610099-3/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2518902 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2518902. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.