Skip to content

Hill Laboratories

3 N Bacton Hill Rd, Frazer, Pennsylvania, United States

FEI 2510425

FDA has 4 inspections on file for Hill Laboratories in Frazer, Pennsylvania (FEI 2510425), from 2015 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 3 citations from this inspection. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2015–2026
0 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.1

      Records of risk management activities in product realization not maintained

      Records of risk management activities in product realization were not maintained.

    2. ISO 13485:2016 Clause 8.2.2

      Complaint handling records not maintained

      Complaint handling records were not adequately maintained.

    3. ISO 13485:2016 Clause 6.2(e)

      Records of personnel competence not maintained

      Appropriate records of personnel education, training and experience were not maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    3. DMR - not or inadequately maintained

      A device master record has not been maintained.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    Show 6 more citations
    1. Management representative implementing the quality system

      The management representative has not ensured that the quality system requirements are effectively established and maintained.

    2. Quality Audits - defined intervals

      Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

    4. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    5. Equipment control activity documentation

      Equipment calibrations and maintenance activities have not been documented.

    6. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

  3. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Show all 9

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2510425 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2510425. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.