Air Techniques Inc
1295 Walt Whitman Rd, Melville, New York, United States
FEI 2428225
FDA has 6 inspections on file for Air Techniques Inc in Melville, New York (FEI 2428225), from 2010 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2010–2019
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been documented.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been documented, approved, adequately documented, and adequately approved.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Show 9 more citations
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual, maintained in a separate portion of the complaint files, and clearly identified.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Implementing Personnel Procedures, Health, Cleanliness.
Requirements addressing the health, cleanliness, and personal practices of employees have not been implemented.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design validation - documentation
The results of design validation, including the individual(s) performing validation, were not adequately documented in the design history file.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Design output - Lack of or inadequate procedures
Procedures for design output have not been established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Change records, content
Records of changes did not include when the change became effective.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not complete.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.184Device history record3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.120Device labeling2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
Show all 14
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.150Storage1 of 4
- 21 CFR 1002.13Annual reports1 of 4
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2428225, one row per refused product line.
2 refusals on 2026-03-06 · charged REGISTERED on 2 · no FDA sample analysis
FDA firm profile- 2026-03-06
UNIT, OPERATIVE DENTAL76EIA
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of West Coast ImportsSCS-2688698-2/20001/1
- 2026-03-06
UNIT, OPERATIVE DENTAL76EIA
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of West Coast ImportsSCS-2688698-2/10001/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2428225 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2428225. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.