Skip to content

Parker Laboratories, Inc.

286 Eldridge Rd, Fairfield, New Jersey, United States

FEI 2212018

FDA has 6 inspections on file for Parker Laboratories, Inc. in Fairfield, New Jersey (FEI 2212018), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. One of the two FOIA requests for this site asked FDA for the 483 from a later inspection, ending Jul 14, 2023, that FDA's Data Dashboard doesn't list yet.

2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.75(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2009–2023
0 OAI · 4 VAI · 2 NAI
Inspections with 483 observations shown
5 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest FY2023
Warning letters
None on file
Import refusals
2 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Items to cover on annual reviews

      Written procedures are not followed for evaluations done at least annually and including provisions for a review of complaints, recalls, returned or salvaged drug products and investigations conducted for each drug product.

    2. Actual yield, % of theoretical yield

      The batch production and control records do not include a statement of the actual yield and percentage of theoretical yield at appropriate stages of processing for each batch of drug product produced.

    3. Reports of Analysis (Components)

      Reports of analysis from component suppliers are accepted in lieu of testing each component for conformity with all appropriate written specifications, without performing at least one specific identity test on each component.

    4. Lack of written stability program

      There is no written testing program designed to assess the stability characteristics of drug products.

    5. Absence of Written Procedures

      Your firm failed to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a process have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    4. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Specified requirements

      Equipment used in the manufacturing process does not meet specified requirements.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that in-process product is controlled until the required inspection and tests have been completed and approved were not implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Jul 14, 2023

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-7489Form 483Closed
  • Inspection ending May 23, 2023NAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 22, 2024FDA-FOIA-2024-7440Form 483· 1 of 4 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.

Show all 13

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2212018, one row per refused product line.

2 refusals from 2023-06-09 to 2023-06-22 · charged NOT LISTED on 2, FRNMFGREG on 2 · no FDA sample analysis

FDA firm profile
  • 2023-06-22

    MEDIA, ELECTROCONDUCTIVE84GYB

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5032623-1/11/1

  • 2023-06-09

    MEDIA, ELECTROCONDUCTIVE84GYB

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5124286-6/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2212018 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2212018. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.