Welch Allyn Inc (Baxter affiliate)

7865 N 86th St, Milwaukee, Wisconsin, United States

FEI 2183461

FDA has 4 inspections on file for Welch Allyn Inc (Baxter affiliate) in Milwaukee, Wisconsin (FEI 2183461), from 2011 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.

16 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.100(a) and 21 CFR 820.198(c).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2011–2022
0 OAI · 4 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
16 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

    4. Software validation for automated processes

      Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.

    5. Complaints

      Complaint files are not adequately maintained.

    Show 2 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Documentation

      Acceptance activities were not adequately documented and adequately maintained as part of the device history record.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.

    4. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    5. Service reports

      Service reports are not documented and do not include the required information.

    Show 2 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2183461, one row per refused product line.

16 refusals from 2017-06-27 to 2026-03-17 · charged NOT LISTED on 11, FRNMFGREG on 10 · no FDA sample analysis

FDA firm profile
  • 2026-03-17

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Northeast ImportsBUU-1787779-2/10001/1

  • 2025-03-06

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northern Border ImportsSCS-8416058-3/11/1

  • 2024-09-24

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3251221-4/11/1

  • 2021-02-09

    ELECTROCARDIOGRAPH74DPS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports97X-0037283-8/11/1

  • 2020-10-22

    ELECTRODE, ELECTROCARDIOGRAPH74DRX

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Northeast ImportsSCS-2370081-0/11/1

Show 11 more
  • 2020-09-23

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northern Border ImportsSCS-1748743-2/11/1

  • 2020-05-29

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-7011722-0/11/1

  • 2020-03-25

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6160482-2/11/1

  • 2020-03-23

    MONITOR, PHYSIOLOGICAL, PATIENT74MHX

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast Imports799-5765846-9/11/1

  • 2020-01-21

    ELECTROCARDIOGRAPH74DPS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-5008352-5/11/1

  • 2019-12-27

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4659824-8/11/1

  • 2019-09-03

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of West Coast Imports799-1039464-0/51/1

  • 2019-05-31

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    DV QUALITY 501(c); 801(a)(3) Adulteration

    Division of Northern Border ImportsSCS-4635986-6/11/1

  • 2017-09-08

    ELECTROCARDIOGRAPH74DPS

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports799-0938156-5/11/1

  • 2017-06-27

    ELECTROCARDIOGRAPH74DPS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Northern Border ImportsUPS-2424718-2/11/1

  • 2017-06-27

    ELECTROCARDIOGRAPH74DPS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Northern Border ImportsUPS-2176199-5/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2183461 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2183461. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.