Skip to content

Staar Surgical Company

1911 Walker Ave, Monrovia, California, United States

FEI 2023826

FDA has 7 inspections on file for Staar Surgical Company in Monrovia, California (FEI 2023826), from 2009 to 2018. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

The most recent OAI inspection ended on Feb 4, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2009–2018
3 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
2 of 7
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    4. Process changes - review, evaluation and revalidation

      A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.

    5. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.

    Show 5 more citations
    1. Incorrect translation to production specifications

      The device design was not correctly translated into production specifications.

    2. QSR

      The quality system record has not been adequately maintained

    3. Software validation for automated processes

      Software used as part of production has not been adequately validated for its intended use according to an established protocol.

    4. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been established.

    5. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Analysis of data/reports from data sources

      Not all data and reports from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    3. Complete files maintained

      Complete complaint files are not maintained.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 7 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Other FEIs recorded as Staar Surgical Company

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2023826 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2023826. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.