Cramer Products Inc
153 W Warren St, Gardner, Kansas, United States
FEI 1910082
FDA has 4 inspections on file for Cramer Products Inc in Gardner, Kansas (FEI 1910082), from 2010 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 13 citations from this inspection: 7 under 21 CFR Part 211, the manufacturing rules for finished drugs, 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2010–2017
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Labeling control records including specimens or copies
Batch production and control records do not include complete labeling control records, including specimens or copies of all labeling used for each batch of drug product produced.
Testing and release for distribution
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications and identity and strength of each active ingredient prior to release.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
DMR - not or inadequately maintained
A device master record has not been maintained.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 8 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Testing and standardization of standards et. al.
Laboratory records do not include complete records of any testing and standardization of laboratory reagents and standard solutions.
Identification of Person Performing Review of Lab Records
Laboratory records are deficient in that they do not include the signature of the second person reviewing the record for accuracy.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Absence of Written Procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.
Written calibration / inspection records not kept
Records of the calibration checks of automatic, mechanical or electronic equipment, including computers or related systems are not maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Analysis of data/reports from data sources
Not all from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Control procedures to monitor and validate performance
Control procedures are not established which validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Cleaning SOPs/instructions
Procedures for the cleaning and maintenance of equipment are deficient regarding sufficient detail of the methods, equipment, and materials used in the cleaning and maintenance operation, and the methods of disassembly and reassembling equipment as necessary to assure proper cleaning and maintenance.
Device identification, control numbers
The device history record does not include certain device identifications and control numbers used.
Variation in the Amount of Components Used
The master production and control records are deficient in that they lack a justification for the variation in the amount of components used in the preparation of a dosage form.
Show 3 more citations
Control/distribution procedures
Procedures for the control and distribution of finished devices have not been established and implemented to ensure that only devices approved for release are distributed.
Quality policy and objectives
The quality policy and objectives were not complete and implemented.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 211.67Equipment cleaning and maintenance1 of 2
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 2
- 21 CFR 211.100Written procedures; deviations1 of 2
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 2
- 21 CFR 211.165Testing and release for distribution1 of 2
- 21 CFR 211.186Master production and control records1 of 2
- 21 CFR 211.188Batch production and control records1 of 2
Show all 18
- 21 CFR 211.194Laboratory records1 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.160Distribution1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1910082 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1910082. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.