FDA has 6 inspections on file for Teco Diagnostics, Inc. in Anaheim, California (FEI 1832216), from 2010 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Oct 14, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2010–2026
- 2 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Info evaluated to determine if event was reportable
The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
Supplier oversight
The type and extent of control to be exercised over the product was not clearly defined.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Verify, validate change {see also 820.100(a)}
The procedures addressing verification or validation of corrective and preventive actions were not defined.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not complete and implemented.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Nonconforming product, Lack of or inadequate procedures
Procedures were not followed for the control of products that do not conform to specifications.
Lack of or inadequate DHR procedures
Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Show 8 more citations
Nature and details of complaint
Records of complaint investigations do not include the nature and details of the complaint.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not implemented.
Responsibility, authority and independence
Employees who manage, perform, and assess work affecting quality have not been assigned the appropriate responsibility and authority to accomplish their work.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Documentation
Process validation activities and results have not been fully documented.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been documented, completed, and implemented.
Quality System Procedures
Quality system procedures were not complete.
Storage procedures to avoid release of unsuitable product
Procedures for controlling the storage of product in storage areas and stock rooms are not adequate to ensure that no obsolete, rejected, or deteriorated product is distributed.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.60Identification1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
Show all 12
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.150Storage1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1832216 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1832216. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.