ELITechGroup, Inc.
370 W 1700 S, Logan, Utah, United States
FEI 1717966
FDA has 4 inspections on file for ELITechGroup, Inc. in Logan, Utah (FEI 1717966), from 2009 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 803.50(a)(1).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2017
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Equipment suitability & capability
Certain test equipment is not suitable for its intended purposes and capable of producing valid results.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Installer records
The person installing the device did not adequately document the inspection and test results to demonstrate proper installation.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Show 2 more citations
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - review and approval
Design input requirements were not reviewed and approved by designated individual(s).
Records of complaint investigation
Records of complaint investigations do not include required information.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Design review - documentation
The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.
Design output - Lack of or inadequate procedures
Procedures for design output have not been established.
Show 2 more citations
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.75Process validation1 of 4
Show all 10
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.170Installation1 of 4
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1717966, one row per refused product line.
2 refusals from 2017-09-28 to 2021-12-03 · charged NOT LISTED on 1, REGISTERED on 1 · no FDA sample analysis
FDA firm profile- 2021-12-03
DEVICE, AUTOMATED SEDIMENTATION RATE81GKB
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsFR1-0018331-3/101/1
- 2017-09-28
DEVICE, IONTOPHORESIS, SPECIFIC USES89KTB
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Northern Border ImportsAEK-3555907-6/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1717966 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1717966. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.