Wilson-Cook Medical Inc.
4900 Bethania Station Rd, Winston Salem, North Carolina, United States
FEI 1037905DUNS 079720472
FDA has 4 inspections on file for Wilson-Cook Medical Inc. in Winston Salem, North Carolina (FEI 1037905), from 2011 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(b).
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for repacking, through Dec 31, 2026. Its registrant is Wilson-Cook Medical (D-U-N-S 079720472).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2011–2025
- 0 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Production processes
Production processes were not monitored to ensure that a device conforms to its specifications.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Acceptance activities were not documented.
Show 1 more citation
Sampling plans
Sampling plans are not based on valid statistical rationale.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Documentation
Process validation activities have not been adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 3 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
Show all 11
- 21 CFR 820.150Storage1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 820.250Statistical techniques1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
FDA drug establishment registration
FDA registration search- Registered operations
- Repacking
- Registration runs through
- Dec 31, 2026
- Registrant
- Wilson-Cook MedicalD-U-N-S 079720472
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1037905 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1037905. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.