Medtronic Navigation, Inc.

200 Medtronic Dr, Lafayette, Colorado, United States

FEI 1000517638

FDA has 12 inspections on file for Medtronic Navigation, Inc. in Lafayette, Colorado (FEI 1000517638), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Feb 4, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.30(g) and 21 CFR 820.100(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
12, 2009–2025
1 OAI · 3 VAI · 8 NAI
Inspections with 483 observations shown
6 of 12
Form 483s posted by FDA
↗
Requested under FOIA
3 requests, latest Aug 20, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    3. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety ActPosted 483 · 1 citation

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Certified product modification

      A previously certified laser product which has been modified, which modification affects any aspects of the product's performance or intended function(s), has not been recertified

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.

    5. Documentation

      Process validation activities have not been adequately documented.

  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  9. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  11. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. General

      Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been implemented.

    3. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not complete and implemented.

    4. DMR required activities {see also 820.181}

      Finished devices were released for distribution prior to completion of activities required in the Device Master Record.

    5. Design control - no procedures

      Procedures to control the design process of the device were not complete.

    Show 4 more citations
    1. Essential design outputs

      Design outputs that are essential for the proper functioning of the device were not completely identified.

    2. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    3. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

    4. Patient Name or Other Identifier

      An individual medical device manufacturer report submitted per FDA Form 3500A did not contain in Block A the patient name or other identifier.

  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Aug 20, 2024. Each one asked for a Form 483.

  • Inspection ending Feb 9, 2024NAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 20, 2024FDA-FOIA-2024-7402Form 483· 1 of 2 sitesClosed
    2. May 1, 2024FDA-FOIA-2024-3913Form 483Closed
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Dec 18, 2020FDA-FOIA-2020-8820Form 483Merged with another request

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 12 inspections, with how many of them cited each.

  • 21 CFR 820.100Corrective and preventive action5 of 6
  • 21 CFR 820.30Design controls4 of 6
  • 21 CFR 820.75Process validation2 of 6
  • 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 6
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 6
  • 21 CFR 820.50Purchasing controls1 of 6
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
  • 21 CFR 820.90Nonconforming product1 of 6
Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000517638 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000517638. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.