Skip to content

Crystal Care International Inc

15941 N Florida Ave, Lutz, Florida, United States

FEI 1000307612

FDA has one inspection on file for Crystal Care International Inc in Lutz, Florida (FEI 1000307612). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 11 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2008
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Servicing - Lack of or inadequate procedures

      Servicing procedures have not been established.

    2. General

      Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established.

    3. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not established.

    4. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been established to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

    5. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not established.

    Show 6 more citations
    1. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not established.

    2. DMR QA procedures and specifications

      The device master record does not include or refer to the location of all quality assurance procedures and specifications.

    3. Analyzing service report

      Service reports were not analyzed following appropriate statistical methods.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.

    5. Lack of or inadequate procedures for labeling

      Labeling procedures were not established.

    6. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not established.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000307612 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000307612. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.