DMX IMAGING, LLC

4159 Corporate Ct Ste B, Palm Harbor, Florida, United States

FEI 1000221974

FDA has 4 inspections on file for DMX IMAGING, LLC in Palm Harbor, Florida (FEI 1000221974), from 2012 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

The most recent OAI inspection ended on Jun 7, 2012. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.22, 21 CFR 820.80(b) and 21 CFR 820.198(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2012–2017
1 OAI · 0 VAI · 3 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

  2. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    4. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    5. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

  3. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Reports and information documentation

      The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act17 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been adequately established.

    2. Report of assembly submitted within 15 days

      A completed report of assembly regarding certified components installed in a diagnostic x-ray system was not submitted to within 15 days following the completion of the assembly.

    3. Lack of or inadequate In-process acceptance procedures

      Procedures for the control of in-process product have not been adequately established.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 12 more citations
    1. Mandatory warning on control panel

      The required warning statement "Warning: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed" is not legible and accessible to view.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification have not been adequately established.

    4. Servicing - Lack of or inadequate procedures

      Procedures or instructions for verifying that servicing meets specified requirements have not been adequately established.

    5. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    6. Half-value layer--amount of filtration

      With reference to Table 1 appearing in Title 21 CFR 1020.30(m), the half-value layer (HVL) of the useful beam for a given x-ray tube potential was less than the appropriate value shown in table 1.

    7. Documentation

      Acceptance activities were not adequately documented.

    8. Warning signs, labels, instructions

      A product report did not describe for each model all warning signs, labels, and instructions for installation, operation, and use that relate to electronic product radiation safety.

    9. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions and reviews by a designated person was not included in the record.

    10. Failure to submit

      You did not submit an annual report for products requiring one.

    11. Personnel

      Personnel do not have the necessary training to perform their jobs.

    12. Failure to submit

      You did not submit a required supplemental report, prior to the introduction of the product into commerce, for a new or modified model within a model or chassis family.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000221974 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000221974. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.