Diagnostic Consulting Network, LLC
3193 Lionshead Ave Ste 200, Carlsbad, California, United States
FEI 1000221318
FDA has 3 inspections on file for Diagnostic Consulting Network, LLC in Carlsbad, California (FEI 1000221318), from 2019 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2019–2026
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 4.2.4
Document Control Procedure Inadequate
The document control procedure does not adequately define controls needed to review and approve documents for adequacy before use and ensure that current versions of applicable documents are available at point of use.
- ISO 13485:2016 Clause 7.1
Records of risk management activities in product realization not maintained
Records of risk management activities in product realization were not adequately maintained.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Documented evaluation
The evaluation of potential suppliers was not documented.
Show 4 more citations
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a process have not been adequately established.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they are being used.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Mar 13, 2026VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 10, 2026. FDAgent holds no copy that FDA posted.
- May 21, 2026FDA-FOIA-2026-5186Form 483· 1 of 5 sitesRecords released
Inspection ending May 6, 2022NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-95Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.75Process validation1 of 3
Show all 10
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.198Complaint files1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000221318 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000221318. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.