Medical Illumination International, Inc
19749 Dearborn St, Chatsworth, California, United States
FEI 1000125341
FDA has 6 inspections on file for Medical Illumination International, Inc in Chatsworth, California (FEI 1000125341), from 2010 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 803.50(a)(2) and 21 CFR 820.30(e).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2010–2025
- 0 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Mar 18, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities have not been adequately documented.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design review - documentation
The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Show 3 more citations
Design input - documentation
Design input requirements were not adequately documented.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Complaints
Complaint files are not maintained.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - production units
The design was not validated using production units or their equivalents.
Design changes - Lack of or Inadequate Procedures
Procedures were not established for the documentation, review, and approval of design changes before their implementation.
Uniform and timely processing
Complaint handling procedures have not been completed and implemented to ensure that all complaints are processed in a uniform and timely manner.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Mar 18, 2025. It asked for a Form 483.
Inspection ending Jan 29, 2025NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 18, 2025FDA-FOIA-2025-2494Form 483· 1 of 2 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 5
- 21 CFR 820.198Complaint files2 of 5
- 21 CFR 820.20Management responsibility1 of 5
- 21 CFR 820.22Quality audit1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.90Nonconforming product1 of 5
Show all 10
- 21 CFR 820.100Corrective and preventive action1 of 5
- 21 CFR 820.184Device history record1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000125341 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000125341. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.