FDA has 11 inspections on file for Pioneer Surgical Technology Inc. in Marquette, Michigan (FEI 1000115331), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Apr 22, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 11, 2009–2024
- 1 OAI · 4 VAI · 6 NAI
- Inspections with 483 observations shown
- 7 of 11
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 2, 2024
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors have not been established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Show 2 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality System Procedures
Quality system procedures and instructions have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Participation; refusal and discontinuance
The informed consent document did not contain a statement that the subject might discontinue participation at any time without penalty or loss of benefits to which the subject was otherwise entitled.
Significant new findings
The informed consent document did not include a statement that significant new findings developed during the course of the research, which might relate to the subject's willingness to continue participation, would be provided to the subject.
Statement of risks
The informed consent document did not contain a statement that the test article or procedure might involve risks to the subject (or to the embryo or fetus, if the subject is or may become pregnant) that are currently unforeseeable.
Promotion or test marketing
An investigational device was promoted prior to FDA approval/clearance.
Sponsor correspondence records inadequate
Records relating to correspondence with an investigator, including required reports are not all complete and current.
Show 5 more citations
Procedures, identification of those which were experimental
The informed consent document did not contain identification of any procedures which were experimental.
No curriculum vitae in investigator agreement
The signed investigator agreements obtained from each participating investigator do not all include the investigator's curriculum vitae.
Whom to contact
The informed consent document lacked an explanation of whom to contact for answers to pertinent questions about the research and research subjects' rights.
No financial disclosure info in investigator agreement
A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement.
Sponsor provide device information to investigators
Not all investigators participating in the investigation were supplied with copies of the investigational plan.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product evaluation/investigation
The evaluation of nonconforming product has not been documented.
Documentation
Acceptance test results of in-process product were not fully documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 2, 2024. It asked for a Form 483.
Inspection ending Jun 14, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Aug 2, 2024FDA-FOIA-2024-6890Form 483· 1 of 4 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 11 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation3 of 7
- 21 CFR 820.100Corrective and preventive action3 of 7
- 21 CFR 820.25Personnel2 of 7
- 21 CFR 820.30Design controls2 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 7
- 21 CFR 820.90Nonconforming product2 of 7
- 21 CFR 820.198Complaint files2 of 7
- 21 CFR 50.25Elements of informed consent1 of 7
Show all 14
- 21 CFR 812.7Prohibition of promotion and other practices1 of 7
- 21 CFR 812.43Selecting investigators and monitors1 of 7
- 21 CFR 812.140Records1 of 7
- 21 CFR 820.20Management responsibility1 of 7
- 21 CFR 820.50Purchasing controls1 of 7
- 21 CFR 812.45Informing investigators1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000115331 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000115331. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.