Compass Health Brands Corp.
910 Callahan Dr Ste 108, Knoxville, Tennessee, United States
FEI 1000113364
FDA has 2 inspections on file for Compass Health Brands Corp. in Knoxville, Tennessee (FEI 1000113364), from 2009 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 13 citations from this inspection: 12 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2009–2016
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over services and suppliers was not clearly defined.
Design control - no procedures
Procedures for design control have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Quality policy and objectives
The quality policy was/were not established by management with executive responsibility.
Show 8 more citations
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been defined and implemented.
Documentation
Acceptance activities were not documented.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not complete.
Management review - Lack of or inadequate procedures
Procedures for management review are not established.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Show 5 more citations
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not defined.
Lack of or inadequate DHR procedures
Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Evaluation of suppliers, contractors, etc., requirements
Adequate quality requirements that must be met by suppliers were not established, defined, and documented.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.184Device history record2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.30Design controls1 of 2
Show all 11
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.181Device master record1 of 2
Other FEIs recorded as Compass Health Brands Corp.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Compass Health Brands Corp.FEI 3005182235Brookpark, Ohio, United States
Last inspected 2026-02-27VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000113364 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000113364. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.