Theranos, Inc. Form 483, September 2015

FDA issued this Form 483 to Theranos, Inc. in Newark, California (FEI 3006231732) at the end of a device inspection that ended September 16, 2015. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Sep 16, 2015
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 17, 2024
3,045 days after the inspection
Facility
FEI 3006231732
Newark, California

FDA's 11 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  2. Documented evaluation

    The evaluation of potential suppliers was not documented.

  3. Acceptable supplier records, inadequate records

    Records of acceptable suppliers have not been adequately established.

  4. Quality Audit/Reaudit - conducted

    Quality audits have not been performed.

  5. Documentation of software validation

    Software validation activities and results for computers or automated data processing systems used as part of production and the quality system have not been documented.

  6. DHR documentation of label release {see also 820.184}

    The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.

  7. Lack of or inadequate DHR procedures

    Procedures for device history records have not been adequately established.

  8. Devices not listed

    Devices for which listing is required have not been listed.

  9. Complaints

    Complaint files are not adequately maintained.

  10. Documentation

    Corrective and preventive action activities and/or results have not been documented.

  11. Investigation of device failures

    Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.