Skip to content

Theranos, Inc.

7333 Gateway Blvd, Newark, California, United States

FEI 3006231732

FDA has one inspection on file for Theranos, Inc. in Newark, California (FEI 3006231732). It ended on and was classified VAI, voluntary action indicated. FDA has also posted a Form 483 dated Oct 27, 2015 that FDAgent has not matched to an inspection on the Data Dashboard.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 807, the device registration and listing rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2015
Inspections with 483 observations shown
1 of 1
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Documented evaluation

      The evaluation of potential suppliers was not documented.

    3. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    4. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    5. Documentation of software validation

      Software validation activities and results for computers or automated data processing systems used as part of production and the quality system have not been documented.

    Show 6 more citations
    1. DHR documentation of label release {see also 820.184}

      The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.

    2. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    3. Devices not listed

      Devices for which listing is required have not been listed.

    4. Complaints

      Complaint files are not adequately maintained.

    5. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    6. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Other FEIs recorded as Theranos, Inc.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

  • Theranos, Inc.FEI 3010479366Palo Alto, California, United States

    Last inspected 2016-08-18VAI

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006231732 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006231732. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.