Miach Orthopaedics
69 Milk St Ste 100, Westborough, Massachusetts, United States
FEI 3020652309
FDA has 2 inspections on file for Miach Orthopaedics in Westborough, Massachusetts (FEI 3020652309), from 2025 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection. They are listed below. One of the five FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Apr 23, 2025. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2025–2026
- 1 OAI · 0 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 5 requests, latest Sep 12, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.7
Sterilization and sterile barrier system validation procedures not documented
Procedures for the validation of processes for sterilization were not adequately documented.
- ISO 13485:2016 Clause 4.2.4
Document Control Procedure Inadequate
The document control procedure does not adequately define controls needed to review, update as necessary, and re-approve documents.
- ISO 13485:2016 Clause 4.2.3
Medical device file does not contain specific requirements
The medical device file(s) does not include a general description and intended use/purpose of the medical device and labeling.
- ISO 13485:2016 Clause 7.3.9
Procedures to control design and development changes not documented
Procedures to control design and development changes were not adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Documented evaluation
The evaluation of potential suppliers and consultants was not documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Show 6 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Sep 12, 2026. Each one asked for a Form 483.
Inspection ending Jul 24, 2026NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 13, 2026. FDAgent holds no copy that FDA posted.
- Sep 10, 2026FDA-FOIA-2026-8806Form 483· 1 of 5 sitesRecords released
Inspection ending Apr 23, 2025OAI
4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 10, 2025. FDAgent holds no copy that FDA posted.
- Sep 12, 2026FDA-FOIA-2026-8865Form 483 + EIROpen
- Sep 9, 2025FDA-FOIA-2025-10780Form 483· 1 of 5 sitesRecords released
- Aug 26, 2025FDA-FOIA-2025-10189Form 483Records released
- Aug 5, 2025FDA-FOIA-2025-8756Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3020652309 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3020652309. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.