Kelyniam Global, Inc.
97 River Rd, Collinsville, Connecticut, United States
FEI 3009424796
FDA has 6 inspections on file for Kelyniam Global, Inc. in Collinsville, Connecticut (FEI 3009424796), from 2013 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
The most recent OAI inspection ended on May 30, 2017. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2013–2024
- 1 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Feb 10, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Design input - documentation
Design input requirements were not documented.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Show 2 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design control - no procedures
Procedures for design control have not been established.
Document review, approval documentation
The documentation of approval of documents does not include the document approval date and the signature of the approving official.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
Show 2 more citations
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Feb 10, 2025. Each one asked for a Form 483.
Inspection ending Aug 23, 2024VAI
3 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Feb 10, 2025FDA-FOIA-2025-1293Form 483· 1 of 4 sitesClosed
- Nov 1, 2024FDA-FOIA-2024-9723Form 483· 1 of 3 sitesClosed
- Nov 1, 2024FDA-FOIA-2024-9661Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation5 of 6
- 21 CFR 820.30Design controls4 of 6
- 21 CFR 820.70Production and process controls3 of 6
- 21 CFR 820.184Device history record3 of 6
- 21 CFR 820.22Quality audit2 of 6
- 21 CFR 820.40[Reserved]2 of 6
- 21 CFR 820.50Purchasing controls2 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 6
Show all 12
- 21 CFR 820.100Corrective and preventive action2 of 6
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 6
- 21 CFR 820.20Management responsibility1 of 6
- 21 CFR 820.25Personnel1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009424796 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009424796. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.