Toby Orthopaedics, Inc.
6355 Sw 8th St Ofc 101, West Miami, Florida, United States
FEI 3008747271
FDA has 4 inspections on file for Toby Orthopaedics, Inc. in West Miami, Florida (FEI 3008747271), from 2013 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2013–2025
- 0 OAI · 4 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not adequately documented in the design history file.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Quality Audits - defined intervals
Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not adequately documented in the design history file.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Show 1 more citation
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Jul 15, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 22, 2025. FDAgent holds no copy that FDA posted.
- Aug 11, 2025FDA-FOIA-2025-9605Form 483· 1 of 5 sitesRecords released
Inspection ending Apr 28, 2023VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-5115Form 483Closed
- FY2023FDA-FOIA-2023-4249Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008747271 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008747271. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.