Skip to content

ZYTO Technologies, Inc.

1064 S North County Blvd Ste 330, Pleasant Grove, Utah, United States

FEI 3008493238

FDA has 3 inspections on file for ZYTO Technologies, Inc. in Pleasant Grove, Utah (FEI 3008493238), from 2014 to 2022. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH · Division of Medical Device and Radiological Health Operations West).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection: 6 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483, and one more for the 483 from a later inspection, ending Jun 6, 2024, that FDA's Data Dashboard doesn't list yet.

No court action or import refusal is filed under this FEI.

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
3, 2014–2022
1 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Sep 19, 2025
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Records of complaint investigation

      Records of complaint investigations do not include required information.

    2. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.

    4. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 2 more citations
    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Complaints

      Complaint files are not adequately maintained.

    3. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    4. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.

    Show 1 more citation
    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    Show 6 more citations
    1. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Design control - no procedures

      Procedures for design control have not been established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    6. DMR - not or inadequately maintained

      A device master record has not been maintained.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 19, 2025. Each one asked for a Form 483.

  • Inspection ending Jun 6, 2024

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Aug 2, 2024FDA-FOIA-2024-6890Form 483· 1 of 4 sitesClosed
  • Inspection ending Dec 13, 2022OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 6, 2025. FDAgent holds no copy that FDA posted.

    1. Sep 19, 2025FDA-FOIA-2025-11098Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008493238 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008493238. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.