Omega Laser Systems
Unit 2, Deans Hall Business Park; Oak Road, Halstead, United Kingdom
FEI 3008440094
FDA has one inspection on file for Omega Laser Systems in Halstead, United Kingdom (FEI 3008440094). It ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 12 citations from this inspection. They are listed below.
FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2015
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- 89-04Red List
- Matched on FDA’s firm name and street address
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety Act12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Documentation
Acceptance activities were not documented and maintained as part of the device history record.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
Show 7 more citations
Design control - no procedures
Procedures for design control have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been.
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision and provisions for remedial action.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - software validation not performed
Validation of device software was not performed.
Failure to submit
You did not submit an annual report by the September 1 deadline for products requiring one.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 820.200Servicing1 of 1
Show all 9
- 21 CFR 1002.13Annual reports1 of 1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2016-03-07
- FDA’s entry
- Omega Laser Systems · Unit 2, Deans Hall Business Park; Oak Road, Halstead, Essex UNITED KINGDOM
- FDA’s inspection record for FEI 3008440094
- Omega Laser Systems · Unit 2, Deans Hall Business Park; Oak Road, Halstead, United Kingdom
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
13 products
- 73 - - - --Anesthesiology
- 74 - - - --Cardiovascular
- 76 - - - --Dental
Show 10 more products
- 77 - - - --Ear,Nose And Throat
- 78 - - - --Gastroenterological & Urological
- 79 - - - --General & Plastic Surgery
- 80 - - - --General Hospital/Personal Use
- 84 - - - --Neurological
- 85 - - - --Obstetrical & Gynecological
- 86 - - - --Ophthalmic
- 87 - - - --Orthopedic
- 89 - - - --Physical Medicine
- 90 - - - --Radiology
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008440094 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008440094. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.