Skip to content

OralDNA Labs, Inc. A Quest Diagnostics Company

214 Overlook Cir Ste 120, Brentwood, Tennessee, United States

FEI 3008381789

FDA has 2 inspections on file for OralDNA Labs, Inc. A Quest Diagnostics Company in Brentwood, Tennessee (FEI 3008381789), from 2011 to 2012. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Nov 23, 2011. No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2011–2012
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Investigation of device failures

      Complaints involving the possible failure of a device and packaging to meet any of its specifications were not investigated where necessary.

    3. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers, contractors, and consultants have not been adequately established.

    4. Documentation

      Acceptance activities were not documented.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained, and implemented.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    4. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    5. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    Show 3 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 9

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008381789 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008381789. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.