First Medical Source LLC
28581 Springfield Dr, Laguna Niguel, California, United States
FEI 3007008418
FDA has 3 inspections on file for First Medical Source LLC in Laguna Niguel, California (FEI 3007008418), from 2012 to 2017. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2012–2017
- 0 OAI · 0 VAI · 3 NAI
- Inspections with 483 observations shown
- 1 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier notification of changes
There is no agreement with suppliers and contractors to notify you of changes in the product or service.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Files do not contain copies of MDR reports
MDR event files do not contain copies of all MDR forms and other information related to the event that was submitted to FDA.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Show 10 more citations
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
Design output - documentation
Design output was not documented before release.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Design input - documentation
Design input requirements were not documented.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
Complaints
Complaint files are not adequately maintained.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.86Acceptance status1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007008418 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007008418. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.