Augustine Temperature Management, LLC
15305 Minnetonka Blvd, Minnetonka, Minnesota, United States
FEI 3005857264
FDA has 5 inspections on file for Augustine Temperature Management, LLC in Minnetonka, Minnesota (FEI 3005857264), from 2010 to 2024. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 5, 2010–2024
- 2 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Aug 28, 2025
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 2 more citations
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Design review - documentation
The design review results, including the individual(s) performing the review, were not documented in the design history file.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Aug 28, 2025. Each one asked for a Form 483.
Inspection ending Nov 27, 2024OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 11, 2025. FDAgent holds no copy that FDA posted.
- Aug 28, 2025FDA-FOIA-2025-10336Form 483Records released
Inspection ending Sep 19, 2023OAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 24, 2023FDA-FOIA-2023-9379Form 483Closed
- Oct 13, 2023FDA-FOIA-2023-8987Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.198Complaint files1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005857264, one row per refused product line.
1 refusal on 2022-12-02 · charged NO 510(K) on 1 · no FDA sample analysis
FDA firm profile- 2022-12-02
SYSTEM, THERMAL REGULATING74DWJ
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border ImportsAEK-2431235-5/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005857264 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005857264. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.