Beijing Toplaser Technology Co., Ltd.
East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, China
FEI 3005450614
FDA has 2 inspections on file for Beijing Toplaser Technology Co., Ltd. in Beijing, China (FEI 3005450614), from 2011 to 2015. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 9 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
1 import refusal is filed under this FEI, on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2011–2015
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - documentation
The results of design validation, including identification of the design, were not adequately documented in the design history file.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 4 more citations
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Documentation
Process validation activities have not been adequately documented.
Certification label or tag
A certification label or tag is not permanently affixed or inscribed.
ID label or tag
An identification label or tag is not permanently affixed or inscribed.
Show 2 more citations
Failure to submit
You did not submit an annual report for products requiring one.
Class IV "Danger" label
Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
Show all 13
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 1002.13Annual reports1 of 2
- 21 CFR 1010.2Certification1 of 2
- 21 CFR 1010.3Identification1 of 2
- 21 CFR 1040.10Laser products1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005450614, one row per refused product line.
1 refusal on 2017-04-19 · charged DIRECTIONS on 1 · no FDA sample analysis
FDA firm profile- 2017-04-19
GENERAL PURPOSE LASER PRODUCTS, NON-MEDICAL95RDW
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-3150851-5/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2009-09-16
- FDA’s entry
- Beijing Toplaser Technology Co., Ltd. · East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, Beijing CHINA
- FDA’s inspection record for FEI 3005450614
- Beijing Toplaser Technology Co., Ltd. · East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, China
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
14 products
- 73 - - - --Anesthesiology
- 74 - - - --Cardiovascular
- 76 - - - --Dental
Show 11 more products
- 77 - - - --Ear,Nose And Throat
- 78 - - - --Gastroenterological & Urological
- 79 - - - --General & Plastic Surgery
- 79 G - - EXPowered laser surgical instruments
- 80 - - - --General Hospital/Personal Use
- 84 - - - --Neurological
- 85 - - - --Obstetrical & Gynecological
- 86 - - - --Ophthalmic
- 87 - - - --Orthopedic
- 89 - - - --Physical Medicine
- 90 - - - --Radiology
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005450614 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005450614. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.