Skip to content

Beijing Toplaser Technology Co., Ltd.

East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, China

FEI 3005450614

FDA has 2 inspections on file for Beijing Toplaser Technology Co., Ltd. in Beijing, China (FEI 3005450614), from 2011 to 2015. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 9 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

1 import refusal is filed under this FEI, on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2011–2015
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
1 refusal
Import alerts
89-08Red List
Matched on FDA’s firm name and street address

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - documentation

      The results of design validation, including identification of the design, were not adequately documented in the design history file.

    2. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    3. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 4 more citations
    1. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    3. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

    4. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Records of complaint investigation

      Records of complaint investigations do not include required information.

    2. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    3. Documentation

      Process validation activities have not been adequately documented.

    4. Certification label or tag

      A certification label or tag is not permanently affixed or inscribed.

    5. ID label or tag

      An identification label or tag is not permanently affixed or inscribed.

    Show 2 more citations
    1. Failure to submit

      You did not submit an annual report for products requiring one.

    2. Class IV "Danger" label

      Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 13

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005450614, one row per refused product line.

1 refusal on 2017-04-19 · charged DIRECTIONS on 1 · no FDA sample analysis

FDA firm profile
  • 2017-04-19

    GENERAL PURPOSE LASER PRODUCTS, NON-MEDICAL95RDW

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-3150851-5/11/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov

    Date published 2009-09-16

    FDA’s entry
    Beijing Toplaser Technology Co., Ltd. · East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, Beijing CHINA
    FDA’s inspection record for FEI 3005450614
    Beijing Toplaser Technology Co., Ltd. · East 3rd Floor, Building M7, NO. 1 Jiuxianqiao Donglu, Beijing, China

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    14 products

    • 73 - - - --Anesthesiology
    • 74 - - - --Cardiovascular
    • 76 - - - --Dental
    Show 11 more products
    • 77 - - - --Ear,Nose And Throat
    • 78 - - - --Gastroenterological & Urological
    • 79 - - - --General & Plastic Surgery
    • 79 G - - EXPowered laser surgical instruments
    • 80 - - - --General Hospital/Personal Use
    • 84 - - - --Neurological
    • 85 - - - --Obstetrical & Gynecological
    • 86 - - - --Ophthalmic
    • 87 - - - --Orthopedic
    • 89 - - - --Physical Medicine
    • 90 - - - --Radiology

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005450614 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005450614. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.