Skip to content

Spinal Solutions, LLC

26157 Jefferson Ave, Murrieta, California, United States

FEI 3004884943

FDA has 2 inspections on file for Spinal Solutions, LLC in Murrieta, California (FEI 3004884943), from 2011 to 2012. The latest ended on and was classified NAI, no action indicated. FDA has also posted a Form 483 dated Jul 23, 2015 that FDAgent has not matched to an inspection on the Data Dashboard.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.120, 21 CFR 820.181 and 21 CFR 820.50.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
2, 2011–2012
0 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
2 of 2
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been established.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Design control - no procedures

      Procedures for design control have not been established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality policy and objectives

      The quality policy, quality objectives, and were not established by management with executive responsibility.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    3. Distribution records

      Distribution records were not maintained and do not include or refer to the location of required information.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    5. Quality Audit/Reaudit - conducted

      Quality audits and reaudits have not been performed.

    Show 11 more citations
    1. DMR - not or inadequately maintained

      A device master record has not been maintained.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    5. Quality System Procedures

      Quality system procedures and instructions have not been established.

    6. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    7. Document review, approval by designated individual

      Documents were not approved by designated individual(s) prior to issuance .

    8. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.

    9. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been established.

    10. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.

    11. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been established.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 13

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004884943 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004884943. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.