Skip to content

Epitope Diagnostics, Inc.

17034 Camino San Bernardo, San Diego, California, United States

FEI 3004185580

FDA has 4 inspections on file for Epitope Diagnostics, Inc. in San Diego, California (FEI 3004185580), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection: 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.198(a) and 21 CFR 820.70(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2009–2023
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Dec 22, 2023
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    3. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Records of complaint investigation

      Records of complaint investigations do not include required information.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Not approved or obsolete document retrieval

      Documents that were not approved and obsolete were observed at a location where they are being used.

    2. Production processes

      Production processes were not controlled to ensure that a device conforms to its specifications.

    3. Storage procedures to prevent mix-ups

      Procedures for controlling the storage of product in storage areas and stock rooms were not established to prevent mix-ups, damage, other adverse effects.

    4. Documentation

      Process validation activities and results have not been fully documented.

    5. Approval of process equipment

      Process equipment has not been approved.

    Show 2 more citations
    1. Calibration standard traceability

      A calibration standard used for measurement equipment is not traceable to national or international standards nor is it an independent reproducible standard.

    2. Process control procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications were not defined and implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Dec 22, 2023. It asked for a Form 483.

  • Inspection ending Apr 28, 2023VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Dec 22, 2023FDA-FOIA-2023-11432Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 10

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004185580 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004185580. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.