Diopsys, Inc.
1000 Kreider Dr, Middletown, Pennsylvania, United States
FEI 3004174467
FDA has 3 inspections on file for Diopsys, Inc. in Middletown, Pennsylvania (FEI 3004174467), from 2010 to 2015. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.181(a) and 21 CFR 820.30(e).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2010–2015
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
DMR device specifications
The device master record does not include or refer to the location of device software specifications.
Design review - documentation
The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Complete files maintained
Complete complaint files are not maintained.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Design changes - Lack of or Inadequate Procedures
Procedures were not completed and followed for the validation or verification of design changes before their implementation.
Design validation - user needs and intended uses
Design validation did not ensure that devices conform to defined user/patient needs and intended uses.
Show 10 more citations
Design review - Lack of or inadequate procedures
Procedures for planning and conducting reviews of the design results at appropriate stages of the device's design development were not complete and implemented.
Processing MDRs {see also 803, 804}
Complaint handling procedures have not been completed and implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups, and reviews by a designated person was not included in the record.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not complete and implemented.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established.
Lack of procedures, or not maintained
Document control procedures were not implemented.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Lack of or inadequate organizational structure
The organization structure is not adequate to assure that quality system requirements are fully met.
Documentation
Process validation activities and results have not been documented.
DMR device specifications
The device master record does not include or refer to the location of all specifications.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.181Device master record2 of 2
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.70Production and process controls1 of 2
Show all 11
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004174467 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004174467. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.