FDA has 6 inspections on file for TEI Biosciences, Inc. in Boston, Massachusetts (FEI 3004170064), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH · Division of Medical Device and Radiological Health Division I).

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Four of the five FOIA requests for this site have asked FDA for this inspection's records.

The most recent OAI inspection ended on Nov 2, 2018. No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.75(a).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2009–2023
1 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
3 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
5 requests, latest FY2023
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, no action was taken to evaluate whether there was any adverse effect on the device's quality.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    5. Software validation for automated processes

      Software used as part of production has not been adequately validated for its intended use according to an established protocol.

    Show 2 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Examination for accuracy

      Labeling was not sufficiently examined by a designated individual for accuracy including the correct expiration date before release.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    4. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    Show 1 more citation
    1. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Five requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.

  • Inspection ending May 17, 2023VAI

    4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 13, 2026. FDAgent holds no copy that FDA posted.

    1. FY2023FDA-FOIA-2023-5777Form 483Closed
    2. Apr 10, 2026FDA-FOIA-2026-3626Form 483Records released
    3. Apr 3, 2024FDA-FOIA-2024-3109EIRClosed
    4. Jan 25, 2024FDA-FOIA-2024-832Form 483Closed
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2023-4390Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 6 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004170064 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004170064. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.