FDA has 6 inspections on file for Z-Medica, LLC in Wallingford, Connecticut (FEI 3004138549), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Nov 3, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.75(a), 21 CFR 820.75(b) and 21 CFR 820.100(b).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2009–2023
- 1 OAI · 2 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Process validation results have not been adequately documented .
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Show 3 more citations
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Records of complaint investigation
Records of complaint investigations do not include required information.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, the verification or validation of corrective actions, and submission of information on quality problems and corrective and preventive actions for management review.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Investigation of device failures
Complaints involving the possible failure of to meet any of its specifications were not evaluated where necessary.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not complete for monitoring and control of process parameters for validated processes.
Documentation
Corrective and preventive action activities have not been documented, including the verification or validation of corrective actions and implementation of corrective and preventive actions.
Show 1 more citation
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.60Identification1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
Show all 11
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004138549 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004138549. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.