Skip to content

Jeil Medical Corporation

702 Kolon Science Valley 2nd, 55, Digital-ro 34, Guro-gu, Guro-gu, Seoul, Korea (the Republic of)

FEI 3004049923

FDA has one inspection on file for Jeil Medical Corporation in Guro-gu, Seoul, Korea (the Republic of) (FEI 3004049923). It ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 14 citations from this inspection: 13 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

22 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2014
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
22 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    4. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    5. Lack of or inadequate In-process acceptance procedures

      Procedures for the control of in-process product have not been adequately established.

    Show 9 more citations
    1. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    2. Training records

      Personnel training is not documented.

    3. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    4. Design validation - documentation

      The results of design validation, including identification of the design, the date, and the individual(s) performing validation, were not adequately documented in the design history file.

    5. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.

    6. DHR documentation of label release {see also 820.184}

      The DHR does not include complete records of examination and release of device labeling, including date and signature of the examiner.

    7. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    8. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

    9. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Show all 11

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004049923, one row per refused product line.

22 refusals from 2018-05-30 to 2025-10-09 · charged REGISTERED on 22, NOT LISTED on 2 · no FDA sample analysis

FDA firm profile
  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/9

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/8

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/7

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/6

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/5

Show 17 more
  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/31

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/30

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/29

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/28

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/27

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/26

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/25

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/23

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/22

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/21

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/20

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/15

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/14

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/11

  • 2025-10-09

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-3112064-8/10002/10

  • 2022-12-09

    ORTHOPEDIC MANUAL SURGICAL INSTRUMENT87LXH

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-1501631-5/11/1

  • 2018-05-30

    HANDPIECE, ROTARY BONE CUTTING76KMW

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports799-6141076-5/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004049923 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004049923. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.