Skip to content

Mi Gwang Contact Lens Co., Ltd.

693 Namcheon-Ro, Namcheon-Myeon, Gyeongsan, Korea (the Republic of)

FEI 3003499948

FDA has 4 inspections on file for Mi Gwang Contact Lens Co., Ltd. in Gyeongsan, Korea (the Republic of) (FEI 3003499948), from 2012 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jul 5, 2012. FDA issued a warning letter on (CDRH). 3 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2012–2022
1 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Dec 17, 2025
Import refusals
3 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Labeling inspection - UDI

      Labeling was not sufficiently examined by a designated individual for accuracy including unique device identifier (UDI) or universal product code (UPC).

    2. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Buildings

      Buildings do not contain sufficient space to perform necessary operations and prevent mix-ups.

    5. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.

    Show 1 more citation
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production processes

      Production processes were not monitored to ensure that a device conforms to its specifications.

    2. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    4. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    Show 1 more citation
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Dec 17, 2025. It asked for a Form 483.

  • Inspection ending Dec 15, 2022VAI

    1 request. In the inspection history above.

    1. Dec 17, 2025FDA-FOIA-2025-14148Form 483· 1 of 2 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Show all 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3003499948, one row per refused product line.

3 refusals from 2018-12-13 to 2022-04-01 · charged NOT LISTED on 3, FRNMFGREG on 3 · no FDA sample analysis

FDA firm profile
  • 2022-04-01

    LENS, CONTACT (DISPOSABLE)86MVN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-7733294-8/21/1

  • 2019-11-21

    LENSES, SOFT CONTACT, DAILY WEAR86LPL

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4143178-3/11/1

  • 2018-12-13

    LENSES, SOFT CONTACT, DAILY WEAR86LPL

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    RXPERSONAL 502(a), 502(f)(1), 801(a)(3), MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast ImportsSCS-2074172-5/11/1B

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003499948 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003499948. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.