Bauer & Haselbarth - Chirurg GmbH
Sauerbruchstr. 7, Ellerau, Germany
FEI 3002806450
FDA has 5 inspections on file for Bauer & Haselbarth - Chirurg GmbH in Ellerau, Germany (FEI 3002806450), from 2010 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 5 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Sep 16, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2010–2026
- 1 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jul 24, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.6
Process for production and service provision not validated
Processes for production and service provision where the resulting output cannot be, or is not, verified by subsequent monitoring or measurement were not adequately validated.
- ISO 13485:2016 Clause 7.4.2
Purchasing information lacks written agreement regarding notification of changes
Purchasing information did not include a written agreement that the supplier notify your firm of changes in purchased product prior to implementation of any changes that affect the ability of purchased product to meet specified purchasing requirements.
- ISO 13485:2016 Clause 8.2.4
Objectivity and impartiality of internal audit process not ensured
The selection of auditors and conduct of audits did not adequately ensure objectivity and impartiality of the audit.
- ISO 13485:2016 Clause 8.3.1
Nonconforming product not adequately evaluated
Nonconforming product was not adequately evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.
- ISO 13485:2016 Clause 4.2.4
Document Control Procedure Inadequate
The document control procedure does not adequately define controls needed to review and approve documents for adequacy before use.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Production processes
Production processes were not conducted to ensure that a device conforms to its specifications.
Show 2 more citations
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jul 24, 2026. It asked for a Form 483.
Inspection ending Mar 19, 2026NAI
1 request. In the inspection history above.
- Jul 24, 2026FDA-FOIA-2026-7133Form 483· 1 of 2 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
Show all 10
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002806450 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002806450. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.