FEI 3002023342

FDA has 5 inspections on file for IVD Research, Inc. in Carlsbad, California (FEI 3002023342), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Oct 9, 2008. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2008–2025
1 OAI · 0 VAI · 4 NAI
Inspections with 483 observations shown
1 of 5
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

    3. ID label, labeling

      The device history record does not include the primary identification label and labeling for each device.

    4. Process control procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established and implemented.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    Show 11 more citations
    1. Calibration documentation

      There is no documentation of calibration dates, the individual performing each calibration, and the next calibration date for measurement and test equipment .

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

    3. Design control - no procedures

      Procedures to control the design process of the device were not complete and implemented.

    4. DMR device specifications

      The device master record does not include or refer to the location of specifications.

    5. Storage procedures to prevent mix-ups

      Procedures for controlling the storage of product in storage areas and stock rooms were not established to prevent mix-ups, damage, other adverse effects.

    6. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers were not defined and documented.

    7. Approval, inadequate purchasing data

      Purchasing information that clearly includes/describes the specified requirements for purchased or otherwise received products and services was not established.

    8. Lack of or inadequate DHR procedures

      Procedures have not been defined to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    9. Uniform and timely processing

      Complaint handling procedures have not been implemented to ensure that all complaints are processed in a uniform and timely manner.

    10. Establish the design and development plan

      A design and development plan has not been established.

    11. Acceptable supplier records

      Records of acceptable suppliers and consultants were not maintained.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 5 inspections.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002023342 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002023342. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.