SC Johnson Professional USA, Inc.
1100 S Highway 27, Stanley, North Carolina, United States
FEI 3001606885
FDA has 6 inspections on file for SC Johnson Professional USA, Inc. in Stanley, North Carolina (FEI 3001606885), from 2010 to 2021. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Oct 30, 2018. FDA issued a warning letter on (CDER). No court action or import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2010–2021
- 1 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDER ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complete test data included in records
Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards.
Deviations from laboratory control requirements
Deviations from written test procedures and laboratory mechanisms are not recorded and justified.
Equipment Design, Size and Location
Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design to facilitate operations for its intended use.
Microbiological Contamination Exam
Each lot of a component that is liable to microbiological contamination that is objectionable in view of its intended use is not subjected to microbiological tests before use.
Testing and release for distribution
Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications prior to release.
Show 4 more citations
Inadequate number of personnel
The number of qualified personnel is inadequate to supervise the manufacture of each drug product.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Written records kept in individual logs
Written records of major equipment use are not included in individual equipment logs.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.
Signatures and dates--person who performs test
Laboratory records do not include the initials or signature of the person who performs each test.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
No written record of investigation
Written records are not always made of investigations into the failure of a batch or any of its components to meet specifications.
Approval and review of procedures
Written procedures are not reviewed and approved by the quality control unit.
Show 5 more citations
Testing Calculations
Laboratory records are deficient in that they do not include all calculations performed during testing.
Component identification test
Specific identification tests are not conducted on components that have been accepted based on the supplier's report of analysis.
Complete Test Data
Laboratory records are deficient in that they do not include a complete record of all data obtained during testing.
Written procedures not established/followed
Written procedures are not followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
In-Process and Laboratory Control Results
The batch production and control records are deficient in that they do not include laboratory control results.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Reprocessing procedures not written or followed
Procedures prescribing a system for reprocessing batches to insure that the reprocessed batches will conform with all established standards, specifications, and characteristics are not written.
Calibration - at intervals, written program, remedial action
The calibration of instruments is not done at suitable intervals in accordance with an established written program.
Identification of Components and In-Process Materials
The batch production and control records are deficient in that they do not include specific identification of each batch of component used.
Signatures and dates--person who performs test
Laboratory records do not include the initials or signature of the person who performs each test.
Backup file not maintained
Failure to maintain a backup file of data entered into the computer or related system.
Show 10 more citations
Component identification test
Specific identification tests are not conducted on components that have been accepted based on the supplier's report of analysis.
Complete Test Data
Laboratory records are deficient in that they do not include a complete record of all data obtained during testing.
Cleaning/maintenance records not kept
Records are not kept for the cleaning, sanitizing and inspection of equipment.
Procedure Deviations Recorded and Justified
Deviations from written production and process control procedures are not recorded.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate designed to assure that conform to appropriate standards of identity, strength, quality and purity.
GMP Training Frequency
GMP training is not conducted with sufficient frequency to assure that employees remain familiar with CGMP requirements applicable to them.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Deviations from laboratory control requirements
Deviations from written specifications, standards, sampling plans, test procedures and laboratory mechanisms are not recorded and justified.
Identity Testing of Each Component
The identity of each component of a drug product is not verified by conducting at least one test to verify the identity, using specific identity tests if they exist.
Written procedures not established/followed
Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Items to cover on annual reviews
Written procedures are not followed for evaluations done at least annually and including provisions for a review of complaints and investigations conducted for each drug product.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions.
Failing drug products not rejected
Drug products failing to meet established specifications are not rejected.
Annual visual exams of drug products
Reserve samples from representative sample lots or batches of drug products selected by acceptable statistical procedures are not examined visually at least once a year for evidence of deterioration.
Calculations performed are in the records
Laboratory records do not include a record of all calculations performed in connection with the test.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 211.194Laboratory records4 of 5
- 21 CFR 211.22Responsibilities of quality control unit3 of 5
- 21 CFR 211.67Equipment cleaning and maintenance3 of 5
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures3 of 5
- 21 CFR 211.100Written procedures; deviations3 of 5
- 21 CFR 211.160General requirements3 of 5
- 21 CFR 211.25Personnel qualifications2 of 5
- 21 CFR 211.165Testing and release for distribution2 of 5
Show all 16
- 21 CFR 211.188Batch production and control records2 of 5
- 21 CFR 211.63Equipment design, size, and location1 of 5
- 21 CFR 211.68Automatic, mechanical, and electronic equipment1 of 5
- 21 CFR 211.115Reprocessing1 of 5
- 21 CFR 211.170Reserve samples1 of 5
- 21 CFR 211.180General requirements1 of 5
- 21 CFR 211.182Equipment cleaning and use log1 of 5
- 21 CFR 211.192Production record review1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001606885 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001606885. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.