FDA has 9 inspections on file for Swiss American CDMO, LLC in Carrollton, Texas (FEI 3001451826), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.75(a) and 21 CFR 820.70(i).

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, through Dec 31, 2027. Its registrant is Swiss-American CDMO, LLC (D-U-N-S 080170933).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
9, 2009–2024
0 OAI · 3 VAI · 6 NAI
Inspections with 483 observations shown
4 of 9
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Feb 19, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Software validation for automated processes

      Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Rejected Material Area

      Separate or defined areas to prevent contamination or mix-ups are deficient regarding operations related to the holding of rejected components and drug product containers before disposition.

    2. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

    3. Complaint Handling Procedure

      Procedures describing the handling of written and oral complaints related to drug products are deficiently written or followed.

    4. GMP Training Frequency

      GMP training is not conducted on a continuing basis to assure that employees remain familiar with CGMP requirements applicable to them.

    5. Storage under appropriate conditions

      Drug products are not stored under appropriate conditions of temperature and humidity so that their identity, strength, quality, and purity are not affected.

    Show 3 more citations
    1. Subdivided component container identification

      Containers holding subdivided components for drug product manufacturing are deficiently identified in that they lack the component name and item code.

    2. Written procedures fail to include

      Written procedures for cleaning and maintenance fail to include description in sufficient detail of methods, equipment and materials used, description in sufficient detail of the methods of disassembling and reassembling equipment as necessary to assure proper cleaning and maintenance and parameters relevant to the operation.

    3. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

  4. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Software validation for automated processes

      Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.

  6. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  7. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  9. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 19, 2025. Each one asked for a Form 483.

  • Inspection ending Dec 16, 2024VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Feb 19, 2025FDA-FOIA-2025-1536Form 483· 1 of 4 sitesClosed
    2. Jan 22, 2025FDA-FOIA-2025-559Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 9 inspections, with how many of them cited each.

Show all 13

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing
Registration runs through
Dec 31, 2027
Registrant
Swiss-American CDMO, LLCD-U-N-S 080170933

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001451826 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001451826. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.